Analysis of Retinol (Vitamin A) using HPLC-UV

Sample:
- Retinoic Acid (68 µg/mL)
- Retinol (55 µg/mL)
HPLC Analytical Conditions: |
|
---|---|
Column: | Develosil HSR C18, 5 µm ( 4.6 x 150 mm) |
Mobile Phase: | Acetonitrile/Water=95/5 |
Flow Rate: | 0.5 mL/min |
Temperature: | 40°C |
Detection: | UV at 325 nm |
Injection Volume: | 2.0 µL |